Close
You have no items in your shopping cart.

Fluke 2640A Data Acquisition Unit

Manufacturer part number: 2640A
Fluke 2640A Data Acquisition Unit, NetDAQ
PURCHASE
AvailableQTYPrice
New:   0
+ -
$0.00
Used:   0
+ -
$0.00Configurations
X

Available Configurations

OPTIONS
NamePurchase
PARALLEL-TO-LAN ADAPTER$0.00
(OBSOLETE, USE 911)$0.00
NETDAQ LOGGER FOR WINDOW$0.00
NETDAQ LOGGER FOR WINDOW W/TRENDING$0.00

The 2640A NetDAQ Networked Data Acquisition Unit is a powerful combination of hardware and software seamlessly integrated to deliver your data directly over your network. These systems, along with optional Trend Link software, enable multiple users to view just the information they need in real time, from anywhere on the system. View current, temperature, voltage, and more on the same screen at the same time. You can also monitor several units simultaneously making it ideal for small-to-medium sized equipment monitoring, product testing, and process validation applications. A NetDAQ unit can also be used as a portable dedicated system connected to a notebook computer for maintenance, product validation, research, and troubleshooting applications.

Combine from one to twenty NetDAQs into an integrated system of up to 400 channels. Use an existing network or simply connect directly to your PC. Two models offer a choice of scan speeds (up to 1,000 rps), and accuracy (up to 0.01%) to meet your needs. The NetDAQ uses Fluke’s patented Universal Input Module which accepts any combination of analog input types for each of its 20 channels - without requiring external signal conditioning

Features:

  • 100, 50, 6 readings/second
  • 18 bit A/D resolves 0.3 µV and .02 °C
  • 300V maximum measurement input
  • Built-in signal conditioning
  • Real time on-board clock
  • -20 to 60 °C (-4 to 140 °F) operating temperature
  • Channel capacity: Analog inputs: 20; Computed channels: 10; Digital I/O and alarm outputs: 8 total; Totalizer: 1
  • Channel Capacity: 20
  • I/O Lines Total: 12
  • Power: 107 to 246V ac (no switching required), 50 and 60 Hz, 15VA maximum; 9V dc to 16V dc, 6W maximum
  • Warm-up Time: 1 hour to rated specifications -or- 15 minutes if relative humidity (non-condensing) is 50% or less
  • Vibration: 0.7g at 15 Hz; 1.3g at 25 Hz; 3g at 55 Hz
  • Shock: 30g half sine per Mil-T-28800E; Bench handling per Mil-T-28800E
  • Operating Temperature: -20° C to 60° C (-4° F to 140° F)
  • Storage Temperature:  -40° C to 70° C (-40° F to 158° F)
  • Relative Humidity: 90% maximum for -10°C to 28°C (14°F to 82.4°F); 75% maximum for 28°C to 35°C (82.4°F to 95°F); 50% maximum for 35°C to 60°C (95°F to 140°F) (3 MΩ range, reduce humidity rating by 25% for 1 hour warm-up. 3 MΩ range meets full humidity ratings with 2 hour warm-up.)
  • Altitude
    • Operating: 2,000 m (6,561 ft) maximum
    • Non-Operating: 12,200 m (40,000 ft) maximum
  • Size:  9.3 cm high, 21.6 cm wide, 36.2 cm deep (3.67 in high, 8.5 in wide, 14.28 in deep)
  • Weight
    • Net: 4 kg (8.8 lb.)
    • Shipping: 6.0 kg (13.2 lb.)
Description

The 2640A NetDAQ Networked Data Acquisition Unit is a powerful combination of hardware and software seamlessly integrated to deliver your data directly over your network. These systems, along with optional Trend Link software, enable multiple users to view just the information they need in real time, from anywhere on the system. View current, temperature, voltage, and more on the same screen at the same time. You can also monitor several units simultaneously making it ideal for small-to-medium sized equipment monitoring, product testing, and process validation applications. A NetDAQ unit can also be used as a portable dedicated system connected to a notebook computer for maintenance, product validation, research, and troubleshooting applications.

Combine from one to twenty NetDAQs into an integrated system of up to 400 channels. Use an existing network or simply connect directly to your PC. Two models offer a choice of scan speeds (up to 1,000 rps), and accuracy (up to 0.01%) to meet your needs. The NetDAQ uses Fluke’s patented Universal Input Module which accepts any combination of analog input types for each of its 20 channels - without requiring external signal conditioning

Features:

  • 100, 50, 6 readings/second
  • 18 bit A/D resolves 0.3 µV and .02 °C
  • 300V maximum measurement input
  • Built-in signal conditioning
  • Real time on-board clock
  • -20 to 60 °C (-4 to 140 °F) operating temperature
  • Channel capacity: Analog inputs: 20; Computed channels: 10; Digital I/O and alarm outputs: 8 total; Totalizer: 1