General Specifications
- User test storage: unlimited
- Remote interface: Ethernet, RJ45
- External memory: USB for USB memory stick
- Display: 7in, 800x480, 24-bit, touch-screen
- AUX interface: D-sub 37p for external CDN, external transformer etc
- External trigger input: 5V TTL
- Synch input: BNC, 10V to 264VAC
- Trigger output: 5V TTL
- External start/stop input: 5V TTL, starts/stops predefined test sequence
- EUT failed input: 5V TTL
- Analog output: 0 to 10V, for use with external options
- Safety circuit: stops the test when unlocked
- Control power: 85VAC to 264VAC, 50/60Hz
- Dimensions, (W x H x D): 45 x 18 x 49cm; 19in/4U
- Weight: 25kg
IEC/EN 61000-4-4 EDITION 2 & 3 EFT/BURST
- Output voltage: 0.2 to 5.0kV ±10% at coaxial output
- Spike frequency: 1Hz to 1MHz
- Polarity: pos/neg/alternate
- Burst duration: 10µs to 1s
- Output impedance: 50Ω
- Burst period: 1ms to 10s
- Rise time: 5ns ±30%
- Test time: 1s to 1000 minutes
- Impulse duration at 50Ω: 50ns ±30%
- Impulse duration at 1000Ω:50ns -15 +100ns
- Trigger: automatic, manual, external trigger input
- Burst mode: normal, continuous, real, random
- Integrated single-phase coupling/decoupling network: 264VAC/16A; 220VDC/10A
IEC/EN 61000-4-5 EDITION 2 SURGE COMBINATION WAVE
- Output voltage 0.2 to 5.0kV ±10%
- Output current: 0.1 to 2.5kA ±10%
- Voltage rise time: 1.2µs ±30%
- Current rise time: 8µs ±20%
- Voltage duration: 50µs ±20%
- Current duration: 20µs ±20%
- Polarity: pos/neg/alternate
- Output impedance: 2Ω
- Integrated single phase CDN: 264VAC/16A; 220VDC/10A
- Phase sync: 0 to 359° with 1° steps or asynchronous mode
- Impulse trigger: automatic 2s to 100 min, manual, external trigger input
- Counter preselect: 1 to 1000/infinite
- Counter: 100000
- Peak voltage monitor: BNC output, 1000:1; display, 3 digits
- Peak current monitor: BNC output, 1kA/V; display, 3 digits
IEC/EN 61000-4-11 EDITION 2 AND IEC/EN 61000-4-29 DIPS & INTERRUPTS
- Max. voltage: 264VAC/DC
- Interrupt time: 0.5 period to 800 periods, 100µs to 1000 minutes
- Max. current: 16A AC/DC continuous
- Max. current for 5s: 20A
- Max. current for 3s: 23A
- Max. current for 3s:40A >500A inrush current
- Interval time 1: period to 800 periods synch; 100µs to 1000 minutes asynch
- Trigger: automatic, manual, external trigger input
- Test time: 1s to 1000 minutes, infinite
- Interrupt dip level: 0%, 0% to 99% with external voltage source
- Phase sync: 0 to 359° 16/40/50/60Hz asynchronous mode