Model: CDT-240 (8 A base, optional 15 A upgrade)
-
Voltage Range: 90–270 Vrms (±30 to 388 V peak)
-
Test Current: 8 A standard, 15 A optional via kit 00-CDT-15A
-
Probe Specs: >100 MΩ, <25 pF via 100:1 BNC output (±2%)
-
Test Time Range: 0.5–10.0 seconds (±1% for ≥1 s; ±10 ms for <1 s)
-
Voltage Accuracy: ±1.5% full scale; ±2% ±0.8 V up to 381 V peak
-
AC Frequency: 50–60 Hz
-
Input Power: 110–240 V AC; fuse 2 A time-delay 5×20 mm
-
Safety: 4 mm CAT III jacks; interlock; survives 6-inch drop
-
Environment: 15–40 °C; 0–90% RH non-condensing
-
Dimensions/Weight: 9¼" × 4¾" × 10"; 5.8 lbs
-
Accessories: Leads, manual, NIST-traceable calibration certificate; optional A2LA
-
Warranty: 1 year
-
Standards Presets: IEC 61010, 60065, 60950, 60601-1, 60335; editable memory