Model: HT‑2000S
Output Voltage (AC): 0–2000 VAC
Maximum Leakage Current: 20 mA
Leakage Current Range: 1–20 mA, adjustable in 0.1 mA steps
Ground Continuity Threshold: Adjustable from 0.08 Ω to 1.50 Ω
Ground Continuity Feature: Auto-offset for test lead resistance, defeatable
Arc Detection: High-frequency independent circuit (50–500 kHz)
Ramp Time: Adjustable from 1 to 5 seconds
Test Duration: Adjustable from 1 to 60 seconds
Voltage Discharge Time: Less than 7 milliseconds
Output Connectors: Front-panel NEMA 5‑15R and 4 mm return socket
Display: 4-digit LED display with ramp bar graph
Indicators: Pass/fail LEDs and audible buzzer
Operation: One-button logic-controlled testing
Dimensions: 9.25″ × 4.75″ × 10″
Weight: Approximately 11 lbs
Power Input: 120 VAC, 60 Hz
Calibration: NIST-traceable certificate included
Compliance: UL and CSA listed; expandable for TUV testing