Close
You have no items in your shopping cart.

Compliance West USA MegaPulse D5-PF Defibrillation Proof Tester

Manufacturer part number: D5-PF
The Compliance West MegaPulse D5-PF Defibrillation Proof Tester is a surge tester built for various requirements.
PURCHASE
Available QTY Price
New:   0
+ -
$0.00
Used:   0
+ -
$0.00 Configurations
X

Available Configurations

OPTIONS
Name Purchase
Gigavac Relay $0.00
Signal Generator $0.00
Continuous Full Voltage $0.00
HV jumper for alternate resistor $0.00
220V only operation $0.00
230V only operation $0.00
240V only operation $0.00
Blood pressure test $0.00
Custom Pelican case for MegaPulse PF $0.00

Surge tester built to the requirements of IEC 60601-1:2005 (Issue 3) Figure 10 (IEC 60601-1 Figure 50 in older versions). Conducts IEC 60601-1:2005 Figure 11; IEC 60601-2-34 Issue 2(Optional); IEC60601- 2-34 Issue 3(Included); IEC 60601-2-49; EC-13; EC-53 and the new requirement of IEC 60601-2- 25:2011 and IEC 60601-2-27:2011 including Correction 1. Tests of protection against the effects of defibrillation (Common mode, Differential mode), Energy reduction test. All these tests use the same 5000V source with difference in output wave shaping. The various wave shaping output circuits required for each test are included in the D5-PF.

Features

  • Output voltage selected digitally on the front panel
  • Integrated High Voltage on hold circuit
  • Integrated Energy measurement
  • Joules readout for the last pulse is displayed by pressing a button
  • USB TestMinder, computer control
  • Y1-Y2 measuring circuit integrated
  • Built-in Pass/Fail reference
  • External interlock disables HV output and defeats test when circuit is open
  • Long life capacitor rated for 2.5 million pulses
  • Front panel indicates internal high voltage and polarity
  • Polarity controlled manually on the front panel
  • Front panel switches and instructions for various test setups
  • Auto-trigger via USB
  • Microcontroller technologies
  • Cables, manual and calibration certificate included
  • One year calibration cycle & one year warranty

The more remarkable changes of the D5-PF version are: output voltage selected digitally, direct joules readout, increased accuracy, faster time between pulses "20 seconds", increased signal generator voltage, pass/fail reference is integrated, dry cap is used and USB Testminder is standard.

    • Charge Voltage: 0 - 5100 V tolerance ±1%
    • Main Capacitance: 32 uF ±5%
    • Inductances: 500uH and 25 mH ±5%
    • Main Resistances: 100 Ω, 50 Ω and 400 Ω ±1% non inductive
    • Voltage Control: Manually, by turning a knob in the front panel or via USB
    • Polarity Control: Positive and Negative. Alternating control available with USB
    • Voltage Display: 4 Digit LED Display
    • Voltage Meter resolution: 5V
    • Voltage Meter Accuracy: ±1%

Duty Cycle:

      • 20s between pulses, ±1% 100 resistor bank, up to 70 pulses
      • 12s between pulses, +1% <-5% 100 resistor bank, continuous duty

Line Voltage: 120V AC, 50/60Hz *optional different line voltages available

Energy Measurement

      • Works only when: 25mH inductor selected (in accordance with the energy reduction test)
      • Display Resolution: 1 Joule
      • Repeatability: ± 1 Joule between pulses
  • Calibration with certificate
  • Manual
  • Test cables
  • Power cord
  • All standard accessories
Description

Surge tester built to the requirements of IEC 60601-1:2005 (Issue 3) Figure 10 (IEC 60601-1 Figure 50 in older versions). Conducts IEC 60601-1:2005 Figure 11; IEC 60601-2-34 Issue 2(Optional); IEC60601- 2-34 Issue 3(Included); IEC 60601-2-49; EC-13; EC-53 and the new requirement of IEC 60601-2- 25:2011 and IEC 60601-2-27:2011 including Correction 1. Tests of protection against the effects of defibrillation (Common mode, Differential mode), Energy reduction test. All these tests use the same 5000V source with difference in output wave shaping. The various wave shaping output circuits required for each test are included in the D5-PF.

Features

  • Output voltage selected digitally on the front panel
  • Integrated High Voltage on hold circuit
  • Integrated Energy measurement
  • Joules readout for the last pulse is displayed by pressing a button
  • USB TestMinder, computer control
  • Y1-Y2 measuring circuit integrated
  • Built-in Pass/Fail reference
  • External interlock disables HV output and defeats test when circuit is open
  • Long life capacitor rated for 2.5 million pulses
  • Front panel indicates internal high voltage and polarity
  • Polarity controlled manually on the front panel
  • Front panel switches and instructions for various test setups
  • Auto-trigger via USB
  • Microcontroller technologies
  • Cables, manual and calibration certificate included
  • One year calibration cycle & one year warranty

The more remarkable changes of the D5-PF version are: output voltage selected digitally, direct joules readout, increased accuracy, faster time between pulses "20 seconds", increased signal generator voltage, pass/fail reference is integrated, dry cap is used and USB Testminder is standard.