Standards Met | FCC 97-270 (part 68); IEC 60255-22-5; IEC 61000-4-11; IEC 61000-4-29; IEC 61000-4-4; IEC 61000-4-5; IEC 61000-4-8; IEC 61000-4-9; IEC 61326; IEC 61850-3 |
Application/Test Type | CE Marking - Cover transient and power fail requirements according to international standards for immunity tests |
Pulse voltage (open circuit) | 160V - 5,000V ± 10% |
Impedance | 2 Ohm |
Combination wave pulse 1, 2/50 - 8/20 μs (Hybrid-Surge pulse) Pulse conforms to IEC/EN 61000-4-5 |
Parameter | Value |
Pulse current (short circuit) | Max. 2,500A ± 10% |
Coupling | Line to line; Line(s) to ground |
Repetition time | 10ms - 9,999ms |
Pulse repetition | max. 1Hz (1s - 999s) |
Phase synchronization | 0° - 360°, resolution 1° (16 - 500Hz) |
Polarity | positive / negative / alternating |
Phase synchronization | 0° - 360°, resolution 1° |
Coupling | L, N, PE; all combinations |
Dips, Interrupts & Variations Conforms to IEC/EN 61000-4-11, IEC/EN 61000-4-29 |
Parameter | Value |
Dips, Interrupts & Variations | As per IEC/EN 61000-4-11, IEC/EN 61000-4-29 and IEC/EN 61000-6-1, -6-2 |
Test duration | 1 - 30,000 or endless, selectable |
Uvar with step transformer | 300V AC + DC |
Test duration | 0:01min - 99:59min |
Peak inrush current capability | 500 Amps |
Burst (EFT) 5/50 ns Pulse conforms to IEC/EN 61000-4-4 |
Parameter | Value |
Pulse amplitude | 200V - 5,500V ± 10%; 100V - 2,750V ± 10% into 50ohm |
Burst frequency | 0.1 kHz - 1,000 kHz |
Polarity | positive / negative |
Uvar with optional variac | N/A |
Switching times | Under 5us into a 100ohm resistive load |
Event time | 20us - 9,999s |
Repetition time | 10ms - 9,999s |
Phase synchronization | 0° - 360°, resolution 1° (16 - 500Hz) |
Burst time | 0.10ms - 999ms |
Dimensions | 19", 3HU |
Weight | 25 kg's (55 Lbs) |