Contact us 800.813.8448

Recently added item(s) ×

You have no items in your shopping cart.

Haefely PEFT 8010 EFT/BURST Generator, 7.3 kV

Model Number: PEFT 8010

View Price / Get Quote

Haefely PEFT 8010 EFT/BURST Generator, 7.3 kV

Click on above image to view full picture

Condition

New

Availability
  • Immediate Purchase
  • Quote to Rent/Lease
  • Quote to Purchase
Description

Haefely PEFT 8010 EFT/BURST Generator, up to 7.3 kV

The PEFT 8010 is the state-of-the-art EFT/Burst Immunity Test System with a built-in coupling/decoupling network for single phase or three-phase mains supply lines and a HV terminal to connect a capacitive coupling clamp IP4B. The spike frequency can be varied from as low as 1Hz to 110kHz.

The PEFT 8010 is capable to generate 999 spikes per millisecond or up to 250 spikes per burst. Its performance fully complies with IEC 61000-4-4:2004, IEC 61000-4-4 Amd.1 Ed.2:01/2010 and IEC 61000-4-4 Ed.3: 03/2012. Pre-programmed Standard Test routines and a number of selected User Test routines make the PEFT 8010 an optimum standalone solution for both testing during development and final testing of products and sub-assemblies.

Test Standards
  •  IEC 61000-4-4
  •   IEC 61000-6-1
  •   IEC 61000-6-2
  •   IEC 61326
  •   IEC 61850-3
Specifications
Wave Shape into 50 ohm load
  • Rise time: 5 ns ± 30 %
  • Impulse duration: 50 ns ± 30 %
Wave Shape into 1000 ohm load
  • Rise time: 5 ns ± 30 %
  • Impulse duration: 50 ns – 15 / + 100 ns
Burst Characteristic
  • Voltage at HV-output: 1.0 … 7.3 kV
  • Voltage at CDN output: 1.0 … 6.6 kV
  • Spike frequency: 1 Hz … 110 kHz
  • Burst duration: 0.01 … 999 ms
  • Burst period: 0.1 Hz … 400 Hz, 2.5 ms … 10 s, 1 period … 500 periods
  • Impulses per second: max. 250 for Unom > 4 kV, max. 500 for Unom ≤ 4 kV
  • Impulses per burst: max. 75 for Unom > 4 kV, max. 150 for Unom ≤ 4 kV
  • Spike distribution normal, continuous, random, real
  • Polarity: positive, negative
  • Test time: 10 s … 8 h per path
  • Single Phase Coupling/Decoupling Network
    • Maximum AC voltage: 264 V @ 16 … 440 Hz
    • Maximum AC current: 16 A @ 50 … 60 Hz, 10 A @ 400 Hz
    • Maximum DC voltage: 125 V
    • Coupling Modes: L-GND, N-GND, PE-GND, LN-GND, LPE-GND, NPEGND, LNPE-GND
    • Residual voltage at Test Supply Input: ≤ 10% of applied test voltage
    Synchronization
    • Frequency: 16 2/3 Hz, 40 Hz, 50 Hz, 60 Hz, 400 Hz, auto
    • Impulse trigger: automatic, manual, external
    System Configuration
    • EUT fail input: BNC, logic low to trigger
    • P90 interface: for a 3-phase CDN extension
    • PESD interface: for connecting an ESD generator
    • Printer interface: Centronix
    • Computer interface: RS-232, IEEE-488 (optional)
    Power Supply
    • Voltage: 85 … 264 VAC, 150 VA
    • Frequency: 50 / 60 Hz
Standard Accessories Included
  • Cable Set
  • User Manual
  • Calibration Certificate
$21,725.20
$21,725.20

Purchase

Request a Quote

Available Options