Teseq CBA 1G-070B Broadband Power Amplifier

New, high power density, design has allowed us to produce a 70 W 1 MHz to 1 GHz amplifier in a compact 3U package. With a minimum of 55 W of linear power this amplifier is ideal for RF immunity testing in a GTEM and for 3V/m testing with a wide range of available antenna.
PURCHASE
AvailableQTYPrice
New:   0
+ -
$0.00
Used:   0
+ -
$0.00Configurations
X

Available Configurations

Teseq CBA 1G-070B Broadband Power Amplifier

New, high power density, design has allowed us to produce a 70 W 1 MHz to 1 GHz amplifier in a compact 3U package. With a minimum of 55 W of linear power this amplifier is ideal for RF immunity testing in a GTEM and for 3V/m testing with a wide range of available antenna.

These amplifiers are easily integrated with the Teseq immunity instruments such as the NSG4070 or ITS 6006 and Teseq GTEM cells to create turnkey IEC61000-4-3 test systems.

The new touch screen colour display gives an immediate visual indication of forward and reverse power along with the current operating status of the amplifier and access to diagnostic information such as gate current and heatsink temperature.

  • Standards
    • IEC/EN 61000-4-3
    • IEC/EN 61000-4-20
    • IEC/EN 61000-4-21
    • CISPR 24
    • EN 61000-6-1
    • EN 61000-6-2
    • IEC 60533
    • IEC 60601-1-2
    • IEC 61326
    • MIL STD 461
    • RTCA/DO 160
    • ISO 11452-2
    • ISO 11452-3
    • ISO 11452-5
    • ISO 11452-11
    • SAE J1113-24
    • SAE J1113-27
  • Harmonics at 55 W: -20 dBc
  • Maximum Input Without Damage: +20 dBm
  • Spurious: -50 dBc
  • Stability: Unconditionally stable
  • Load VSWR Without Damage (All Phase Angles): Infinite
  • Load VSWR Without Foldback (All Phase Angles): 4:1
  • RF Input: Type N (socket)
  • RF Output: Type N (socket)
  • Gain Control: 0-30 dB in 255 steps
  • Output Impedance: 50 Ω
  • Input VSWR: 2:1
  • Output VSWR: 2:1
  • Interface: Touch screen, GPIB, RS232, Ethernet & USB
  • Safety Interlock: BNC Connector, Open for Standby
  • Supply Voltage: 90 - 264 VAC
  • Supply Frequency: 47 - 63 Hz
  • Supply Power Requirement: M0.4 kVa
  • Dimensions: 3U, 19 inch rack width 510 mm deep
  • Weight: 16 kg
  • Temperature Operational: 0°C – 40°C
  • Temperature Non-Operating: -40°C – 85°C
  • Humidity (Non-Condensing): 95%
Description

Teseq CBA 1G-070B Broadband Power Amplifier

New, high power density, design has allowed us to produce a 70 W 1 MHz to 1 GHz amplifier in a compact 3U package. With a minimum of 55 W of linear power this amplifier is ideal for RF immunity testing in a GTEM and for 3V/m testing with a wide range of available antenna.

These amplifiers are easily integrated with the Teseq immunity instruments such as the NSG4070 or ITS 6006 and Teseq GTEM cells to create turnkey IEC61000-4-3 test systems.

The new touch screen colour display gives an immediate visual indication of forward and reverse power along with the current operating status of the amplifier and access to diagnostic information such as gate current and heatsink temperature.

  • Standards
    • IEC/EN 61000-4-3
    • IEC/EN 61000-4-20
    • IEC/EN 61000-4-21
    • CISPR 24
    • EN 61000-6-1
    • EN 61000-6-2
    • IEC 60533
    • IEC 60601-1-2
    • IEC 61326
    • MIL STD 461
    • RTCA/DO 160
    • ISO 11452-2
    • ISO 11452-3
    • ISO 11452-5
    • ISO 11452-11
    • SAE J1113-24
    • SAE J1113-27